SMS Settings TRAY.SIZE=50TRAY.NUMBER=2100 (for 50 samples)TRAY.NUMBER=2050 (for 100 samples) CP31:RACK => 1-50CP32:RACK => 51-100 Typical vertical is 19.2 above rack while training. To format a 720KB floppy +> format a: /T:80 /N:9
Settings for dof during pwxcal
To add the display of Protune during tuning type the following. create(‘showprotunegui’,’string’,’global’) showprotunegui=’y’
NMR Reference Standards from Millipore Sigma
Adjusting Mass Flow on a Varian Cryo Platform. Gas Regulator Tuning Procedure (MK1) With the CCC cold and operating at equilibrium (running at least 3 hours), the gas regulator valve should quickly be closed to finger tight and then opened 5 turns. The vernier needle valve on the inlet side of the cold loop should…
Calibrate the Mass Flow Meter and Pressure Transducers. 1. Follow the procedures in “Calibration of Mass Flow Meter and Pressure Transducers” 2. Continue with “Adjust the Mass Flow and Ballast Tank Valves,” Adjust the Mass Flow and Ballast Tank Valves Procedure for reduction of vibrations associated with the compressed gas. 1. Remove the housing surrounding the CCC….
Cryo-driver DC condition: 1. Put system into tune mode. 2. Remove cable from pre amp input J5331, and install a BNC “T” in place. 3. Connect one end of “T” to voltmeter to measure Voltage, leave other end of “T” open. Measure voltage at J5331 (voltage should = 5.0V +/- .1V). 4. Connect 50 ohm load on “T”, measure…
BBOD When VnmrJ starts and the status says “Initializing” or VnmrJ does not start the problem is some inconsistency in the network. VnmrJ adm will start even if VnmrJ does not, but will also come up with the black bar of death “Initializing”. In /etc/hosts, check that the hostname and the IP address are correct….