SMS – Reverse Trays
If you need to reverse the trays, CP31 for CP32, you must also train the rack position CP31:CLEAR. (9) Define Rack CP31:CLEAR (M)ove
If you need to reverse the trays, CP31 for CP32, you must also train the rack position CP31:CLEAR. (9) Define Rack CP31:CLEAR (M)ove
Consoles with a PPC may not boot up because the bootfile in /etc/bootptab is VxWorks, and not vxWorksPPC. Change this. Also, change the bootfile in /vnmr/acq/bootptab to vxWorksPPC or a subsequent setacq will overwrite the bootfile name. The PPC should have the bootfile (vxWorksPPC) in nvram, and on boot, the console requests the bootfile. The…
VnmrJ Tips and Tricks
SMS Settings TRAY.SIZE=50TRAY.NUMBER=2100 (for 50 samples)TRAY.NUMBER=2050 (for 100 samples) CP31:RACK => 1-50CP32:RACK => 51-100 Typical vertical is 19.2 above rack while training. To format a 720KB floppy +> format a: /T:80 /N:9
Settings for dof during pwxcal
To add the display of Protune during tuning type the following. create(‘showprotunegui’,’string’,’global’) showprotunegui=’y’
NMR Reference Standards from Millipore Sigma
Adjusting Mass Flow on a Varian Cryo Platform. Gas Regulator Tuning Procedure (MK1) With the CCC cold and operating at equilibrium (running at least 3 hours), the gas regulator valve should quickly be closed to finger tight and then opened 5 turns. The vernier needle valve on the inlet side of the cold loop should…
Calibrate the Mass Flow Meter and Pressure Transducers. 1. Follow the procedures in “Calibration of Mass Flow Meter and Pressure Transducers” 2. Continue with “Adjust the Mass Flow and Ballast Tank Valves,” Adjust the Mass Flow and Ballast Tank Valves Procedure for reduction of vibrations associated with the compressed gas. 1. Remove the housing surrounding the CCC….