Cold Probe Preamp Testing

Cryo-driver DC condition:

1.       Put system into tune mode.

2.       Remove cable from pre amp input J5331, and install a BNC “T” in place.

3.       Connect one end of “T” to voltmeter to measure Voltage, leave other end of “T” open. Measure voltage at J5331 (voltage should = 5.0V +/- .1V).

4.       Connect 50 ohm load on “T”, measure voltage, (voltage should = ~1.65V)

Internal Pre-amp DC condition:

1.       Put system into tune mode.

2.       Remove 50 ohm from “T” and install preamp cable P5331.

3.       Measured voltage should now be between 2.50-3.0V (limits 2.25-3.25) with preamp at room temp or between 2.75-3.25V(limits 2.5-3.5) with preamp at operating temperature (25K).

Note: Preamp voltages outside the limits will require factory repair.

Pre-amp Diode test:

1. Use a DVM set to the “diode test” Mode

2. Place the positive meter lead on the center conductor of the H1 transmit port, of the cold probe, and the negative meter lead on the shield portion of the connector

3. Verify the following voltages:

·         Approximately 1.3 VDC when warm (room temperature)

·         Approximately 1.9 VDC when at operating temperature (25 degrees Kelvin)

Pre-amp impedance test:

1. Use a DVM set to the “resistance measurement” mode

2. Measure the impedance at the H1 transmit port of the cold probe

·         You should read approximately 5 to 20 Mega Ohms

3. Measure the impedance at the H1 receive port of the cold probe

·         You should read approximately 30 to 40 Ohms

Similar Posts

  • acq Debug

    The closest thing there is is the np checklist. The one other thing that may give you a clue as to the culprit if the problem is reproducible, is to run debug. You will need to set up mini com. In a terminal window, cd script filename start minicom In the VnmrJ command line, type…

  • Mass Flow Setting

    Calibrate the Mass Flow Meter and Pressure Transducers. 1. Follow the procedures in “Calibration of Mass Flow Meter and Pressure Transducers” 2. Continue with “Adjust the Mass Flow and Ballast Tank Valves,” Adjust the Mass Flow and Ballast Tank Valves Procedure for reduction of vibrations associated with the compressed gas. 1. Remove the housing surrounding the CCC….

  • vxWorks PPC

    Consoles with a PPC may not boot up because the bootfile in /etc/bootptab is VxWorks, and not vxWorksPPC. Change this. Also, change the bootfile in /vnmr/acq/bootptab to vxWorksPPC or a subsequent setacq will overwrite the bootfile name. The PPC should have the bootfile (vxWorksPPC) in nvram, and on boot, the console requests the bootfile. The…

  • BBOD

    BBOD When VnmrJ starts and the status says “Initializing” or VnmrJ does not start the problem is some inconsistency in the network. VnmrJ adm will start even if VnmrJ does not, but will also come up with the black bar of death “Initializing”. In /etc/hosts, check that the hostname and the IP address are correct….