SMS Settings
SMS Settings
TRAY.SIZE=50
TRAY.NUMBER=2100 (for 50 samples)
TRAY.NUMBER=2050 (for 100 samples)
CP31:RACK => 1-50
CP32:RACK => 51-100
Typical vertical is 19.2 above rack while training.
To format a 720KB floppy +> format a: /T:80 /N:9
Settings for dof during pwxcal
Cryo-driver DC condition: 1. Put system into tune mode. 2. Remove cable from pre amp input J5331, and install a BNC “T” in place. 3. Connect one end of “T” to voltmeter to measure Voltage, leave other end of “T” open. Measure voltage at J5331 (voltage should = 5.0V +/- .1V). 4. Connect 50 ohm load on “T”, measure…
If you need to reverse the trays, CP31 for CP32, you must also train the rack position CP31:CLEAR. (9) Define Rack CP31:CLEAR (M)ove
Consoles with a PPC may not boot up because the bootfile in /etc/bootptab is VxWorks, and not vxWorksPPC. Change this. Also, change the bootfile in /vnmr/acq/bootptab to vxWorksPPC or a subsequent setacq will overwrite the bootfile name. The PPC should have the bootfile (vxWorksPPC) in nvram, and on boot, the console requests the bootfile. The…