SMS Settings
SMS Settings
TRAY.SIZE=50
TRAY.NUMBER=2100 (for 50 samples)
TRAY.NUMBER=2050 (for 100 samples)
CP31:RACK => 1-50
CP32:RACK => 51-100
Typical vertical is 19.2 above rack while training.
To format a 720KB floppy +> format a: /T:80 /N:9
VnmrJ Tips and Tricks
BBOD When VnmrJ starts and the status says “Initializing” or VnmrJ does not start the problem is some inconsistency in the network. VnmrJ adm will start even if VnmrJ does not, but will also come up with the black bar of death “Initializing”. In /etc/hosts, check that the hostname and the IP address are correct….
To add the display of Protune during tuning type the following. create(‘showprotunegui’,’string’,’global’) showprotunegui=’y’
The closest thing there is is the np checklist. The one other thing that may give you a clue as to the culprit if the problem is reproducible, is to run debug. You will need to set up mini com. In a terminal window, cd script filename start minicom In the VnmrJ command line, type…
Adjusting Mass Flow on a Varian Cryo Platform. Gas Regulator Tuning Procedure (MK1) With the CCC cold and operating at equilibrium (running at least 3 hours), the gas regulator valve should quickly be closed to finger tight and then opened 5 turns. The vernier needle valve on the inlet side of the cold loop should…