UMKC – 2H Signal

Customer reported no 2H signal during find lock function or gradient shimming. Other issues, AMT had bad fan and the console main switch

  • would not stay on.
  • Replaced fans in AMT
  • Replaced main power switch
  • Tested system RF

When testing RF found that output of AMT had good power out for H1 but the RF power out for H2 and C13 has low. AMT should provide about

60 dB of gain and we found 8 dB of gain. This resulted in inadequate power out to get usable pulses for H2 or C13.

Also noted that the good H1 signal had 120 Hz ripple modulated on it. The 120 Hz source was traced to the PTS 500.

Recommend the Following:

  • Send AMT to our repair depot for refurbishment. This service will provide a full rebuild of the AMT and comes with 1 one year warranty. Lead time about 4 weeks
  • Send in PTS for repair by OEM. If available, this will be the least expansive route. If the repair cannot be done, we have spare PTS 500’s in stock.

We will reach out to PTS and request for quotes. Once we have all of the quotes, I will prepare an estimate and send it.

PTS output with 120 Hz modulation.

Similar Posts

  • Concordia PM_080921

    Probe HCN_0775_sem HCN_3512_sem Shims HCN_0775_sem_lk_d20 HCN_3512_sem_lk_d2o h1pw90 7.8us @ 52 7.5us @ 51 lshp_ns 0.74/8.59/10.6 0.60/5.18/8.09 lshp 0.33/2.73/4.92 0.30/1.89/2.09 h1sn 702:1 877:1 f19pw90 9us @ 52 — f19sn 109:1 — c13pwx 14.8us @ 60 24.7us @ 60 c13gamah2 8065 @ 43 8065 @ 48 n15pwx 33us @ 60 — gzprofile 32.8 G/cm 62.8 G/cm gzrecovery…

  • Carbonet Move

    De-Energize and Pack De-energized and packed system for shipment to Canada. System was de-energized and warmed up and then packed into Oxford shipping crate (including sides). Four pieces were shipped to IET for final crating and packing. Customer arranged for pickup by a freight carrier. Found the following issued during benchmark; Remote Status box inop…

  • Arrowhead – 2H Signal

    Customer reported that they could not lock (using 2H) or gradient shim. Looked at 2H RF path and good signal coming out of console to probe. Problem must be on receive side. Swapped the following parts for troubleshooting: 2H Diplexer Lock Preamp 2H RF Relay (verified operation) 2H Filter (bypassed) Believe issue is the Mixer…

  • NAU – Mercury

    Mercury 400 Alarm Found that the PS204, which is a +12VDC and +5VDC power supply had a capacitor that was bulging.  Tested the +12VDC and observed the following: When OK 12.4 VDC 0.1 VAC When Beeping 11.8 VDC 0.8 VAC At Startup 10.5 VDC 1.5V AC Ordered replacement power supply from Mouser Electronics.  Part Number: HCC512-A+G…