UMKC – 2H Signal

Customer reported no 2H signal during find lock function or gradient shimming. Other issues, AMT had bad fan and the console main switch

  • would not stay on.
  • Replaced fans in AMT
  • Replaced main power switch
  • Tested system RF

When testing RF found that output of AMT had good power out for H1 but the RF power out for H2 and C13 has low. AMT should provide about

60 dB of gain and we found 8 dB of gain. This resulted in inadequate power out to get usable pulses for H2 or C13.

Also noted that the good H1 signal had 120 Hz ripple modulated on it. The 120 Hz source was traced to the PTS 500.

Recommend the Following:

  • Send AMT to our repair depot for refurbishment. This service will provide a full rebuild of the AMT and comes with 1 one year warranty. Lead time about 4 weeks
  • Send in PTS for repair by OEM. If available, this will be the least expansive route. If the repair cannot be done, we have spare PTS 500’s in stock.

We will reach out to PTS and request for quotes. Once we have all of the quotes, I will prepare an estimate and send it.

PTS output with 120 Hz modulation.

Similar Posts

  • Odyssey – SMS Install

    Completed Install of SMS Robot and SYS-V Controller. Customer had table, cables and interface cards. Vertical Reach Rotation CP1:AT.HAND.GP 1.8 30.3 89.9 CP1:CLEAR.HAND.GP 24.8 0.0 89.9 HAND.GP 16.2 19.9 89.8 CP31:RACK P1 19.5 4.7 76.3 CP31:RACK P2 19.5 4.6 43.2 CP31:RACK P3 19.5 27.2 49.3 CP191:MAGNET 19.3 21.9 272.0 CP191:CLEAR 34.0 0.0 272.0

  • Arrowhead – 2H Signal

    Customer reported that they could not lock (using 2H) or gradient shim. Looked at 2H RF path and good signal coming out of console to probe. Problem must be on receive side. Swapped the following parts for troubleshooting: 2H Diplexer Lock Preamp 2H RF Relay (verified operation) 2H Filter (bypassed) Believe issue is the Mixer…

  • Carbonet Move

    De-Energize and Pack De-energized and packed system for shipment to Canada. System was de-energized and warmed up and then packed into Oxford shipping crate (including sides). Four pieces were shipped to IET for final crating and packing. Customer arranged for pickup by a freight carrier. Found the following issued during benchmark; Remote Status box inop…

  • RFUMS – 070324

    System had several issues including bad Digital PS, inconsistent shims and intermittent RF output. Believe the shim issues were due to bad digital PS. Still having intermittent RF issues. Also, the CPU board was bricked. I was able to restore it to defaults using procedure below. Attach a terminal connection to the Acq CPU Diag…