NAU – Mercury
Mercury 400 Alarm
Found that the PS204, which is a +12VDC and +5VDC power supply had a capacitor that was bulging. Tested the +12VDC and observed the following:
| When OK | 12.4 VDC | 0.1 VAC |
| When Beeping | 11.8 VDC | 0.8 VAC |
| At Startup | 10.5 VDC | 1.5V AC |
Ordered replacement power supply from Mouser Electronics.
Part Number: HCC512-A+G
Also replaced two power supply fans for the Mercury.
Newark part number:
ADDA AA1281HB-AT Axial Fan, Square, High Speed, Ball, 115 VAC, 120 mm, 38 mm, AA12038 Series, 94.5 cu.ft/min


Manufacturer:ADDAManufacturer Part No:AA1281HB-ATNewark Part No.:58P0728Product RangeAA12038 SeriesTechnical
Inova 500 sidebands
Customer reported 120 Hz sidebands. Appears that the PTS620 has power supply issues causing 120 Hz sidebands in spectra. It is more pronounced with channel 1 then channel 2, but it is present in both.
Customer also has a PTS D320 on channel 3 and 4. When I tried to swap that in for testing, the frequency is off by several MHz on both channels. So, PTS D320 is also bad.

EnlargeMouser Part #:675-HCC512A-GManufacturer Part #:
HCC512-A+G
Manufacturer:
Condor / SL Power
Description:Linear Power Supplies 5V@6A 9-15V @ 2.5AHCC512-A+G Datasheet
| Product Category: | Linear Power Supplies | |
| Manufacturer: | SL Power | |
| RoHS: | Details | |
| Product: | Linear Supplies | |
| Output Power: | 30 W | |
| Input Voltage: | 100 VAC/120 VAC/215 VAC/230 VAC/240 VAC | |
| Number of Outputs: | 2 Output | |
| Output Voltage-Channel 1: | 5 VDC | |
| Output Current-Channel 1: | 6 A | |
| Output Voltage-Channel 2: | 9 VDC to 15 VDC Adjustable | |
| Output Current-Channel 2: | 2.5 A | |
| Mounting Style: | Panel | |
| Length: | 9.38 in | |
| Width: | 4.88 in | |
| Height: | 2.75 in | |
| Open Frame/Enclosed: | Open Frame | |
| Commercial/Medical: | Commercial | |
| Brand: | Condor / SL Power | |
| Load Regulation: | Regulated | |
| Series: | Commercial Linear | |
| Factory Pack Quantity: | 5 | |
| Part # Aliases: | HCC512-A+-G |