FSU – Water Damage

Water Damage

NMR had a failure with the air dryer system that caused water to enter the airlines of system all the way thru the shim coil. Had to replace the following:

  • All system air solenoids
  • Shim Coil

Also had to clean the upper barrel for proper spinning.

Shim Coil was replaced with an Oxford 23/54 coil vs. Varian 27/54. This coil is 10 mm longer and has a different 50 pin connector. Needed to build an adapter for cable and use a 10 mm spacer.

Cryo Shim Quench

after repairs and coil replacement, fond that cryo line was greater than 1.2 KHz wide. After reviewing He fill procedure with customer, determined that they must have quenched a cryoshim due to hot gases during the fill.

Trained the customer on proper fill procedure and RRI mapped and cryoshimed magnet. Was able to achieve RT lineshape specs after successful 3D Gradient shimming.

Cryo ShimValue (Amps)
Z1+4.660
Z2+5.080Z2 value may be reversed. See notes in magnet book.
Z3+8.000
X-4.257
Y-11.914
ZX+3.700
ZY+1.600
XY-6.010(S2 on ISS10)
X2-Y2+2.110(C2 on ISS10)

RF Amp Testing

After running pw90 specs, found that both Hi Band and Low Band had low output. Ran RF gain test and found the gain of the High Band AMP was 41 dB (vs 49-53 dB) and the Low Band AMP was 46.4 dB (vs 55-60 dB). Recommend repairing amp modules.

XmtrAttn/SWAMP OutProbe
CH1(H1)7.80.341.140
CH2(C13)9.42.946.444.8

Customer Training

Recommend user operator training by Don Emerson and Don Frank that includes operation, application and technical aspects of the NMR system.

Similar Posts

  • Odyssey – SMS Install

    Completed Install of SMS Robot and SYS-V Controller. Customer had table, cables and interface cards. Vertical Reach Rotation CP1:AT.HAND.GP 1.8 30.3 89.9 CP1:CLEAR.HAND.GP 24.8 0.0 89.9 HAND.GP 16.2 19.9 89.8 CP31:RACK P1 19.5 4.7 76.3 CP31:RACK P2 19.5 4.6 43.2 CP31:RACK P3 19.5 27.2 49.3 CP191:MAGNET 19.3 21.9 272.0 CP191:CLEAR 34.0 0.0 272.0

  • Concordia PM_080921

    Probe HCN_0775_sem HCN_3512_sem Shims HCN_0775_sem_lk_d20 HCN_3512_sem_lk_d2o h1pw90 7.8us @ 52 7.5us @ 51 lshp_ns 0.74/8.59/10.6 0.60/5.18/8.09 lshp 0.33/2.73/4.92 0.30/1.89/2.09 h1sn 702:1 877:1 f19pw90 9us @ 52 — f19sn 109:1 — c13pwx 14.8us @ 60 24.7us @ 60 c13gamah2 8065 @ 43 8065 @ 48 n15pwx 33us @ 60 — gzprofile 32.8 G/cm 62.8 G/cm gzrecovery…

  • NAU – Mercury

    Mercury 400 Alarm Found that the PS204, which is a +12VDC and +5VDC power supply had a capacitor that was bulging.  Tested the +12VDC and observed the following: When OK 12.4 VDC 0.1 VAC When Beeping 11.8 VDC 0.8 VAC At Startup 10.5 VDC 1.5V AC Ordered replacement power supply from Mouser Electronics.  Part Number: HCC512-A+G…

  • RFUMS – 070324

    System had several issues including bad Digital PS, inconsistent shims and intermittent RF output. Believe the shim issues were due to bad digital PS. Still having intermittent RF issues. Also, the CPU board was bricked. I was able to restore it to defaults using procedure below. Attach a terminal connection to the Acq CPU Diag…