SMS – Reverse Trays
If you need to reverse the trays, CP31 for CP32, you must also train the rack position CP31:CLEAR.
(9) Define Rack
CP31:CLEAR
(M)ove
VnmrJ Tips and Tricks
BBOD When VnmrJ starts and the status says “Initializing” or VnmrJ does not start the problem is some inconsistency in the network. VnmrJ adm will start even if VnmrJ does not, but will also come up with the black bar of death “Initializing”. In /etc/hosts, check that the hostname and the IP address are correct….
To add the display of Protune during tuning type the following. create(‘showprotunegui’,’string’,’global’) showprotunegui=’y’
SMS Settings TRAY.SIZE=50TRAY.NUMBER=2100 (for 50 samples)TRAY.NUMBER=2050 (for 100 samples) CP31:RACK => 1-50CP32:RACK => 51-100 Typical vertical is 19.2 above rack while training. To format a 720KB floppy +> format a: /T:80 /N:9
SMS alignment off by about 1/2 cm causing sample move failure. Re-aligned SMS. Ve Re Rt CP1:AT.HAND.GP 2.0 30.4 90.0 CP1:CLEAR.HAND.GP 23.0 0.0 90.0 HAND.GP 16.5 19.9 90.1 CP31:RACK P1 19.5 4.7 76.6 CP31:RACK P2 19.5 4.7 43.7 CP31:RACK P3 19.5 27.2 49.7 CP191:MAGNET 13.4 29.6 259.3 CP191:CLEAR 34.0 0.0 259.3
Calibrate the Mass Flow Meter and Pressure Transducers. 1. Follow the procedures in “Calibration of Mass Flow Meter and Pressure Transducers” 2. Continue with “Adjust the Mass Flow and Ballast Tank Valves,” Adjust the Mass Flow and Ballast Tank Valves Procedure for reduction of vibrations associated with the compressed gas. 1. Remove the housing surrounding the CCC….