Arrowhead – 2H Signal

Customer reported that they could not lock (using 2H) or gradient shim. Looked at 2H RF path and good signal coming out of console to probe. Problem must be on receive side. Swapped the following parts for troubleshooting:

  • 2H Diplexer
  • Lock Preamp
  • 2H RF Relay (verified operation)
  • 2H Filter (bypassed)

Believe issue is the Mixer Module (PN 0191422700)


MR400-RF

Swapped mixer module and problem still persisted. Looked at LO coming into mixer module and found it was the correct frequency and amplitude but when below 80 MHz, it was not stable. When looking at it with a frequency counter, the frequency shifted about 1 KHz. When looking at frequencies about 80 Mhz, the signal was steady. Traced the instability back to the RF coming out of the Synth Board output.

Synth Board is not available in OTI or NMRA stock. Customer will need to order directly from Agilent.

Similar Posts

  • Concordia PM_080921

    Probe HCN_0775_sem HCN_3512_sem Shims HCN_0775_sem_lk_d20 HCN_3512_sem_lk_d2o h1pw90 7.8us @ 52 7.5us @ 51 lshp_ns 0.74/8.59/10.6 0.60/5.18/8.09 lshp 0.33/2.73/4.92 0.30/1.89/2.09 h1sn 702:1 877:1 f19pw90 9us @ 52 — f19sn 109:1 — c13pwx 14.8us @ 60 24.7us @ 60 c13gamah2 8065 @ 43 8065 @ 48 n15pwx 33us @ 60 — gzprofile 32.8 G/cm 62.8 G/cm gzrecovery…

  • RFUMS – 070324

    System had several issues including bad Digital PS, inconsistent shims and intermittent RF output. Believe the shim issues were due to bad digital PS. Still having intermittent RF issues. Also, the CPU board was bricked. I was able to restore it to defaults using procedure below. Attach a terminal connection to the Acq CPU Diag…

  • Carbonet Move

    De-Energize and Pack De-energized and packed system for shipment to Canada. System was de-energized and warmed up and then packed into Oxford shipping crate (including sides). Four pieces were shipped to IET for final crating and packing. Customer arranged for pickup by a freight carrier. Found the following issued during benchmark; Remote Status box inop…