Abbott RF Check – 072321

Checked Advance III HD RF output on H1.

At PLW1 of 100dBmWatts
SGU2 Out1.33 dBm1.4 mw
BLA Out51.78 dBm150.6 w
QNP Out51.33 dBm135.8 w
HPPR In49.48 dBm88.7 w
Probe In49.24 dBm83.9 w
PLW1 at BLA OutdBmWatts
1041.75 dBm14.9 w
2044.66 dBm29.2 w
3046.41 dBm43.7 w
PLW1 at Probe IndbmWatts
1038.71 dBm7.5 w
2042.88 dBm19.4 w
3043.50 dBm22.4 w

Power out at probe appears to be enough to meet pw90 spec on probe.

When plw1 is set to 16 watts we can get a pw90 of 20 us.

When plw1 is set to 30 watts we only get a pw90 of 18.5 us.

Believe that probe may be bad.

Similar Posts

  • NAU – Mercury

    Mercury 400 Alarm Found that the PS204, which is a +12VDC and +5VDC power supply had a capacitor that was bulging.  Tested the +12VDC and observed the following: When OK 12.4 VDC 0.1 VAC When Beeping 11.8 VDC 0.8 VAC At Startup 10.5 VDC 1.5V AC Ordered replacement power supply from Mouser Electronics.  Part Number: HCC512-A+G…

  • Carbonet Move

    De-Energize and Pack De-energized and packed system for shipment to Canada. System was de-energized and warmed up and then packed into Oxford shipping crate (including sides). Four pieces were shipped to IET for final crating and packing. Customer arranged for pickup by a freight carrier. Found the following issued during benchmark; Remote Status box inop…

  • Concordia PM_080921

    Probe HCN_0775_sem HCN_3512_sem Shims HCN_0775_sem_lk_d20 HCN_3512_sem_lk_d2o h1pw90 7.8us @ 52 7.5us @ 51 lshp_ns 0.74/8.59/10.6 0.60/5.18/8.09 lshp 0.33/2.73/4.92 0.30/1.89/2.09 h1sn 702:1 877:1 f19pw90 9us @ 52 — f19sn 109:1 — c13pwx 14.8us @ 60 24.7us @ 60 c13gamah2 8065 @ 43 8065 @ 48 n15pwx 33us @ 60 — gzprofile 32.8 G/cm 62.8 G/cm gzrecovery…