ProTune Display
BydonfrankTo add the display of Protune during tuning type the following. create(‘showprotunegui’,’string’,’global’) showprotunegui=’y’
acq Debug
BydonfrankThe closest thing there is is the np checklist. The one other thing that may give you a clue as to the culprit if the problem is reproducible, is to run debug. You will need to set up mini com. In a terminal window, cd script filename start minicom In the VnmrJ command line, type…
Massflow Adjustment
BydonfrankAdjusting Mass Flow on a Varian Cryo Platform. Gas Regulator Tuning Procedure (MK1) With the CCC cold and operating at equilibrium (running at least 3 hours), the gas regulator valve should quickly be closed to finger tight and then opened 5 turns. The vernier needle valve on the inlet side of the cold loop should…
VnmrJ Tips and Tricks
BydonfrankVnmrJ Tips and Tricks
Settings for dof during pwxcal
BydonfrankSettings for dof during pwxcal
SMS – Reverse Trays
BydonfrankIf you need to reverse the trays, CP31 for CP32, you must also train the rack position CP31:CLEAR. (9) Define Rack CP31:CLEAR (M)ove