SMS – Reverse Trays
If you need to reverse the trays, CP31 for CP32, you must also train the rack position CP31:CLEAR.
(9) Define Rack
CP31:CLEAR
(M)ove
Calibrate the Mass Flow Meter and Pressure Transducers. 1. Follow the procedures in “Calibration of Mass Flow Meter and Pressure Transducers” 2. Continue with “Adjust the Mass Flow and Ballast Tank Valves,” Adjust the Mass Flow and Ballast Tank Valves Procedure for reduction of vibrations associated with the compressed gas. 1. Remove the housing surrounding the CCC….
SMS alignment off by about 1/2 cm causing sample move failure. Re-aligned SMS. Ve Re Rt CP1:AT.HAND.GP 2.0 30.4 90.0 CP1:CLEAR.HAND.GP 23.0 0.0 90.0 HAND.GP 16.5 19.9 90.1 CP31:RACK P1 19.5 4.7 76.6 CP31:RACK P2 19.5 4.7 43.7 CP31:RACK P3 19.5 27.2 49.7 CP191:MAGNET 13.4 29.6 259.3 CP191:CLEAR 34.0 0.0 259.3
Adjusting Mass Flow on a Varian Cryo Platform. Gas Regulator Tuning Procedure (MK1) With the CCC cold and operating at equilibrium (running at least 3 hours), the gas regulator valve should quickly be closed to finger tight and then opened 5 turns. The vernier needle valve on the inlet side of the cold loop should…
The closest thing there is is the np checklist. The one other thing that may give you a clue as to the culprit if the problem is reproducible, is to run debug. You will need to set up mini com. In a terminal window, cd script filename start minicom In the VnmrJ command line, type…
NMR Reference Standards from Millipore Sigma
BBOD When VnmrJ starts and the status says “Initializing” or VnmrJ does not start the problem is some inconsistency in the network. VnmrJ adm will start even if VnmrJ does not, but will also come up with the black bar of death “Initializing”. In /etc/hosts, check that the hostname and the IP address are correct….