Abbott RF Check – 072321

Checked Advance III HD RF output on H1.

At PLW1 of 100dBmWatts
SGU2 Out1.33 dBm1.4 mw
BLA Out51.78 dBm150.6 w
QNP Out51.33 dBm135.8 w
HPPR In49.48 dBm88.7 w
Probe In49.24 dBm83.9 w
PLW1 at BLA OutdBmWatts
1041.75 dBm14.9 w
2044.66 dBm29.2 w
3046.41 dBm43.7 w
PLW1 at Probe IndbmWatts
1038.71 dBm7.5 w
2042.88 dBm19.4 w
3043.50 dBm22.4 w

Power out at probe appears to be enough to meet pw90 spec on probe.

When plw1 is set to 16 watts we can get a pw90 of 20 us.

When plw1 is set to 30 watts we only get a pw90 of 18.5 us.

Believe that probe may be bad.