Carbonet Move

De-Energize and Pack

De-energized and packed system for shipment to Canada. System was de-energized and warmed up and then packed into Oxford shipping crate (including sides). Four pieces were shipped to IET for final crating and packing. Customer arranged for pickup by a freight carrier.

Found the following issued during benchmark;

  • Remote Status box inop
  • Missing Lock reject filter
  • Probe glass insert broken
  • AMT power out on low band is low

Recommended to customer to replace lock reject filter and remote status box and to have probe repaired by JSR. AMT can probably wait as its output is usable. All specs on benchmark passed (other than issued noted above).

H1C13
CH 110 dBm12.8 dBm
CH 29.6 dBm12.2 dBm
AMT46.8 dBm (47 watts)50.7 dBm (120 watts)

Carbonet-Benchmark

Similar Posts

  • UMKC – 2H Signal

    Customer reported no 2H signal during find lock function or gradient shimming. Other issues, AMT had bad fan and the console main switch would not stay on. Replaced fans in AMT Replaced main power switch Tested system RF When testing RF found that output of AMT had good power out for H1 but the RF…

  • FSU Shimming – 100322

    Lineshape does not meet spec. Needs 3D Gradient shimming. Found that the upper barrel was not centered in magnet field. Re-centered upper barrel. Performed 3D gradient shimming successfully. Obtained lineshape spec after RT shimming. Performed H1 and C13 performance specs, all passed. Default shims and power level vales saved. New gradient map saved.

  • FSU – Water Damage

    Water Damage NMR had a failure with the air dryer system that caused water to enter the airlines of system all the way thru the shim coil. Had to replace the following: All system air solenoids Shim Coil Also had to clean the upper barrel for proper spinning. Shim Coil was replaced with an Oxford…

  • NAU – Mercury

    Mercury 400 Alarm Found that the PS204, which is a +12VDC and +5VDC power supply had a capacitor that was bulging.  Tested the +12VDC and observed the following: When OK 12.4 VDC 0.1 VAC When Beeping 11.8 VDC 0.8 VAC At Startup 10.5 VDC 1.5V AC Ordered replacement power supply from Mouser Electronics.  Part Number: HCC512-A+G…