Service Reports

  • FSU – Water Damage

    Water Damage NMR had a failure with the air dryer system that caused water to enter the airlines of system all the way thru the shim coil. Had to replace the following: All system air solenoids Shim Coil Also had to clean the upper barrel for proper spinning. Shim Coil was replaced with an Oxford…

  • Carbonet Move

    De-Energize and Pack De-energized and packed system for shipment to Canada. System was de-energized and warmed up and then packed into Oxford shipping crate (including sides). Four pieces were shipped to IET for final crating and packing. Customer arranged for pickup by a freight carrier. Found the following issued during benchmark; Remote Status box inop…

  • NAU – Mercury

    Mercury 400 Alarm Found that the PS204, which is a +12VDC and +5VDC power supply had a capacitor that was bulging.  Tested the +12VDC and observed the following: When OK 12.4 VDC 0.1 VAC When Beeping 11.8 VDC 0.8 VAC At Startup 10.5 VDC 1.5V AC Ordered replacement power supply from Mouser Electronics.  Part Number: HCC512-A+G…