Loyola He Fill – 080322

Filled three NMR magnets using 100L dewer.

MagnetStartEndTotal
Oxford 50028%100%72%
95L45L50L
8:208:4020 mins
Oxford 300Unknown100%NA
45L10L35L
8:509:0515 mins
Bruker 50081%91%10%
10L0L10L
9:159:205 mins

Similar Posts

  • Arrowhead – 2H Signal

    Customer reported that they could not lock (using 2H) or gradient shim. Looked at 2H RF path and good signal coming out of console to probe. Problem must be on receive side. Swapped the following parts for troubleshooting: 2H Diplexer Lock Preamp 2H RF Relay (verified operation) 2H Filter (bypassed) Believe issue is the Mixer…

  • Carbonet Move

    De-Energize and Pack De-energized and packed system for shipment to Canada. System was de-energized and warmed up and then packed into Oxford shipping crate (including sides). Four pieces were shipped to IET for final crating and packing. Customer arranged for pickup by a freight carrier. Found the following issued during benchmark; Remote Status box inop…

  • NAU – Mercury

    Mercury 400 Alarm Found that the PS204, which is a +12VDC and +5VDC power supply had a capacitor that was bulging.  Tested the +12VDC and observed the following: When OK 12.4 VDC 0.1 VAC When Beeping 11.8 VDC 0.8 VAC At Startup 10.5 VDC 1.5V AC Ordered replacement power supply from Mouser Electronics.  Part Number: HCC512-A+G…

  • FSU Shimming – 100322

    Lineshape does not meet spec. Needs 3D Gradient shimming. Found that the upper barrel was not centered in magnet field. Re-centered upper barrel. Performed 3D gradient shimming successfully. Obtained lineshape spec after RT shimming. Performed H1 and C13 performance specs, all passed. Default shims and power level vales saved. New gradient map saved.