UGA – Doty Probe Install

nstalled DOTY HX probe. Lineshape was marginal, as was S/N. Had to install sample from below like a solids probe. Was unable to get it to work with the upper barrel and sample eject because the probe is longer and has too much air leakage. See message below sent to DOTY.

Judy, I was assisting Audie Roberts at UGA in installing the Doty 5mm H-F, X 5mm PFG probe on his 600. We were able to get the probe installed using the sample holder and inserting it from the bottom of the magnet. But, when we tried to use the Varian upper barrel assy. To insert the sample from the top, we could not develop enough pleasure to eject the sample.

We removed the upper barrel and probe assy. From the magnet and hooked it up to test the eject. We also did this with a Varian probe. This is what we found. The Varian probe still ejected fine. The Doty probe would not. We noticed two things. One, the brass sleeve is about 5mm shorter than the depth of that part of the Varian probe. This means that the upper barrel does not go into the Doty probe as far, which I believe allows more air to escape. Also, and even more dominate, the brass sleeve has a gap between it and the silver probe shell. This allows a significant amount of eject air to escape. We tested this theory by sealing the gap during eject and the sample would eject. It still required more air than the Varian probe.

I have attached a couple of pics. One showing the Varian probe mated with the upper barrel, and the other the Doty probe. Maybe a solution would be to extend the brass sleeve by 5mm and to also fill the gap with RTV? Any other war to reduce the eject air leakage would improve the eject function of the probe. If you have any questions, let Audie or myself know. Thanks.

Similar Posts

  • RFUMS – 070324

    System had several issues including bad Digital PS, inconsistent shims and intermittent RF output. Believe the shim issues were due to bad digital PS. Still having intermittent RF issues. Also, the CPU board was bricked. I was able to restore it to defaults using procedure below. Attach a terminal connection to the Acq CPU Diag…

  • Carbonet Move

    De-Energize and Pack De-energized and packed system for shipment to Canada. System was de-energized and warmed up and then packed into Oxford shipping crate (including sides). Four pieces were shipped to IET for final crating and packing. Customer arranged for pickup by a freight carrier. Found the following issued during benchmark; Remote Status box inop…

  • FSU Shimming – 100322

    Lineshape does not meet spec. Needs 3D Gradient shimming. Found that the upper barrel was not centered in magnet field. Re-centered upper barrel. Performed 3D gradient shimming successfully. Obtained lineshape spec after RT shimming. Performed H1 and C13 performance specs, all passed. Default shims and power level vales saved. New gradient map saved.

  • NAU – Mercury

    Mercury 400 Alarm Found that the PS204, which is a +12VDC and +5VDC power supply had a capacitor that was bulging.  Tested the +12VDC and observed the following: When OK 12.4 VDC 0.1 VAC When Beeping 11.8 VDC 0.8 VAC At Startup 10.5 VDC 1.5V AC Ordered replacement power supply from Mouser Electronics.  Part Number: HCC512-A+G…