SMS Settings
SMS Settings
TRAY.SIZE=50
TRAY.NUMBER=2100 (for 50 samples)
TRAY.NUMBER=2050 (for 100 samples)
CP31:RACK => 1-50
CP32:RACK => 51-100
Typical vertical is 19.2 above rack while training.
To format a 720KB floppy +> format a: /T:80 /N:9
The closest thing there is is the np checklist. The one other thing that may give you a clue as to the culprit if the problem is reproducible, is to run debug. You will need to set up mini com. In a terminal window, cd script filename start minicom In the VnmrJ command line, type…
If you need to reverse the trays, CP31 for CP32, you must also train the rack position CP31:CLEAR. (9) Define Rack CP31:CLEAR (M)ove
Calibrate the Mass Flow Meter and Pressure Transducers. 1. Follow the procedures in “Calibration of Mass Flow Meter and Pressure Transducers” 2. Continue with “Adjust the Mass Flow and Ballast Tank Valves,” Adjust the Mass Flow and Ballast Tank Valves Procedure for reduction of vibrations associated with the compressed gas. 1. Remove the housing surrounding the CCC….
VnmrJ Tips and Tricks
To add the display of Protune during tuning type the following. create(‘showprotunegui’,’string’,’global’) showprotunegui=’y’