SMS – Reverse Trays
If you need to reverse the trays, CP31 for CP32, you must also train the rack position CP31:CLEAR. (9) Define Rack CP31:CLEAR (M)ove
If you need to reverse the trays, CP31 for CP32, you must also train the rack position CP31:CLEAR. (9) Define Rack CP31:CLEAR (M)ove
SMS alignment off by about 1/2 cm causing sample move failure. Re-aligned SMS. Ve Re Rt CP1:AT.HAND.GP 2.0 30.4 90.0 CP1:CLEAR.HAND.GP 23.0 0.0 90.0 HAND.GP 16.5 19.9 90.1 CP31:RACK P1 19.5 4.7 76.6 CP31:RACK P2 19.5 4.7 43.7 CP31:RACK P3 19.5 27.2 49.7 CP191:MAGNET 13.4 29.6 259.3 CP191:CLEAR 34.0 0.0 259.3