Massflow Adjustment

Adjusting Mass Flow on a Varian Cryo Platform.

Gas Regulator Tuning Procedure (MK1)

  • With the CCC cold and operating at equilibrium (running at least 3 hours), the gas regulator valve should quickly be closed to finger tight and then opened 5 turns. The vernier needle valve on the inlet side of the cold loop should also be opened fully. At this point the mass flow through the system will be a maximum. This maximum mass flow, Mmax. is typically between 39 and 45 SLM when the entire system is cold, stable and the compressor is charged to 210 psi. Record Mmax·
  • The vernier needle valve should then be closed until the mass flow becomes stable at: Mmax/2 + 9
  • Finally, close the gas regulator needle valve slowly until the mass flow through the cold loop stabilizes at 18. This may take several minutes to accomplish as the regulator bottle takes time to charge/discharge and reach equilibrium. 

Similar Posts

  • BBOD

    BBOD When VnmrJ starts and the status says “Initializing” or VnmrJ does not start the problem is some inconsistency in the network. VnmrJ adm will start even if VnmrJ does not, but will also come up with the black bar of death “Initializing”. In /etc/hosts, check that the hostname and the IP address are correct….

  • vxWorks PPC

    Consoles with a PPC may not boot up because the bootfile in /etc/bootptab is VxWorks, and not vxWorksPPC. Change this. Also, change the bootfile in /vnmr/acq/bootptab to vxWorksPPC or a subsequent setacq will overwrite the bootfile name. The PPC should have the bootfile (vxWorksPPC) in nvram, and on boot, the console requests the bootfile. The…

  • Mass Flow Setting

    Calibrate the Mass Flow Meter and Pressure Transducers. 1. Follow the procedures in “Calibration of Mass Flow Meter and Pressure Transducers” 2. Continue with “Adjust the Mass Flow and Ballast Tank Valves,” Adjust the Mass Flow and Ballast Tank Valves Procedure for reduction of vibrations associated with the compressed gas. 1. Remove the housing surrounding the CCC….

  • acq Debug

    The closest thing there is is the np checklist. The one other thing that may give you a clue as to the culprit if the problem is reproducible, is to run debug. You will need to set up mini com. In a terminal window, cd script filename start minicom In the VnmrJ command line, type…