Settings for dof during pwxcal
| H1 | 13C | 15N | 31P |
|---|---|---|---|
| 300 | -9000 | -9200 | 7000 |
| 400 | -11000 | -12000 | 9000 |
| 500 | -14500 | 0 | 12000 |
| 600 | -17000 | 700 | 14000 |
| 750 | -21000 | 1100 | 17500 |
| 800 | -22600 | 1200 | 18500 |
| J | 151 | 86 | 20 |
To add the display of Protune during tuning type the following. create(‘showprotunegui’,’string’,’global’) showprotunegui=’y’
Cryo-driver DC condition: 1. Put system into tune mode. 2. Remove cable from pre amp input J5331, and install a BNC “T” in place. 3. Connect one end of “T” to voltmeter to measure Voltage, leave other end of “T” open. Measure voltage at J5331 (voltage should = 5.0V +/- .1V). 4. Connect 50 ohm load on “T”, measure…
BBOD When VnmrJ starts and the status says “Initializing” or VnmrJ does not start the problem is some inconsistency in the network. VnmrJ adm will start even if VnmrJ does not, but will also come up with the black bar of death “Initializing”. In /etc/hosts, check that the hostname and the IP address are correct….
Consoles with a PPC may not boot up because the bootfile in /etc/bootptab is VxWorks, and not vxWorksPPC. Change this. Also, change the bootfile in /vnmr/acq/bootptab to vxWorksPPC or a subsequent setacq will overwrite the bootfile name. The PPC should have the bootfile (vxWorksPPC) in nvram, and on boot, the console requests the bootfile. The…
Adjusting Mass Flow on a Varian Cryo Platform. Gas Regulator Tuning Procedure (MK1) With the CCC cold and operating at equilibrium (running at least 3 hours), the gas regulator valve should quickly be closed to finger tight and then opened 5 turns. The vernier needle valve on the inlet side of the cold loop should…